Epilepsia. 2014 Sep 19. doi: 10.1111/epi.12774. [Epub ahead of print]
Safety of a dedicated brain MRI protocol in patients with a vagus nerve stimulator.
de Jonge JC1, Melis GI, Gebbink TA, de Kort GA, Leijten FS.
Author information
- 1Department of Neurology and Neurosurgery, Brain Center Rudolf Magnus, University Medical Center Utrecht, Utrecht, The Netherlands.
Abstract
Although implanted metallic devices constitute a relative contraindication to magnetic resonance imaging (MRI) scanning, the safety of brain imaging in a patient with a vagus nerve stimulator (VNS) is classified as "conditional," provided that specific manufacturer guidelines are followed when a transmit and receive head coil is used at 1.5 or 3.0 Tesla. The aim of this study was to evaluate the safety of performing brain MRI scans in patients with the VNS. From September 2009 until November 2011, 101 scans were requested in 73 patients with the VNS in The Netherlands. Patients were scanned according to the manufacturer's guidelines. No patient reported any side effect, discomfort, or pain during or after the MRI scan. In one patient, a lead break was detected based on device diagnostics after the MRI-scan. However, because no system diagnostics had been performed prior to MR scanning in this patient, it is unclear whether MR scanning was responsible for the lead break. The indication for most scans was epilepsy related. Twenty-six scans (26%) were part of a (new) presurgical evaluation and could probably better have been performed prior to VNS implantation. Performing brain MRI scans in patients with an implanted VNS is safe when a modified MRI protocol is followed.
Wiley Periodicals, Inc. © 2014 International League Against Epilepsy.
KEYWORDS:
Epilepsy; Magnetic resonance imaging; Patient safety; Vagus nerve stimulator
- PMID:
- 25244102
- [PubMed - as supplied by publisher] http://www.ncbi.nlm.nih.gov/pubmed/25244102
Related citations in PubMed
- 3 Tesla MRI of patients with a vagus nerve stimulator: initial experience using a T/R head coil under controlled conditions.[J Magn Reson Imaging. 2010]
3 Tesla MRI of patients with a vagus nerve stimulator: initial experience using a T/R head coil under controlled conditions.
Gorny KR, Bernstein MA, Watson RE Jr. J Magn Reson Imaging. 2010 Feb; 31(2):475-81.
- MRI of the brain is safe in patients implanted with the vagus nerve stimulator.[Seizure. 2001]
MRI of the brain is safe in patients implanted with the vagus nerve stimulator.
Benbadis SR, Nyhenhuis J, Tatum WO 4th, Murtagh FR, Gieron M, Vale FL. Seizure. 2001 Oct; 10(7):512-5.
- Vagus nerve stimulation therapy system: in vitro evaluation of magnetic resonance imaging-related heating and function at 1.5 and 3 tesla.[Neuromodulation. 2006]
Vagus nerve stimulation therapy system: in vitro evaluation of magnetic resonance imaging-related heating and function at 1.5 and 3 tesla.
Shellock FG, Begnaud J, Inman DM. Neuromodulation. 2006 Jul; 9(3):204-13.
- Review Magnetic resonance imaging compatibility and safety of the SOUNDTEC Direct System.[Laryngoscope. 2006]
Review Magnetic resonance imaging compatibility and safety of the SOUNDTEC Direct System.
Dyer RK Jr, Nakmali D, Dormer KJ. Laryngoscope. 2006 Aug; 116(8):1321-33.
- Review Modern management of epilepsy: Vagus nerve stimulation.[Baillieres Clin Neurol. 1996]
Review Modern management of epilepsy: Vagus nerve stimulation.
Ben-Menachem E. Baillieres Clin Neurol. 1996 Dec; 5(4):841-8.
No comments: